MDL vs MDEL: an MDL licenses a Class II-IV device, while an MDEL applies to an establishment performing specified activities such as importing or distributing. A business may need one, both, or neither, depending on its devices, activities, and exemptions.

This guide is for medical-device manufacturers, importers, and distributors planning to sell devices in Canada or reviewing an existing Canadian supply chain. It focuses on the practical distinction between an MDL and an MDEL, and on the information to confirm before deciding which obligations apply.

What is the difference between an MDL and an MDEL?

The two licences cover different regulatory subjects.

An MDL, or medical device licence, applies to a specific device. Under section 26 of the Medical Devices Regulations, SOR/98-282, a manufacturer must hold an MDL before selling a Class II, III, or IV medical device in Canada.

An MDEL, or medical device establishment licence, concerns an establishment and specified activities, including importing and distributing medical devices. It is not product approval and does not replace an MDL for a Class II-IV device.

A company can have obligations on both sides of the transaction:

  • The device may require an MDL because of its class.
  • The Canadian establishment may require an MDEL because of the activities it performs.
  • An exemption may affect the MDEL assessment.
Licence What it covers Relevant business roles or activities Key qualification
MDL A specific medical device Manufacturer seeking to sell a Class II-IV device in Canada The requirement is based on device class. Class I devices do not require an MDL under section 26.
MDEL An establishment performing specified regulated activities Establishments importing or distributing medical devices Assess the activity and exemptions under the regulations. It is separate from the device’s MDL.
Both may apply The device and the establishment’s activities A business manufacturing a Class II-IV device and also performing activities requiring an MDEL Determine each obligation separately and confirm exemptions against actual activities.

Which licence applies to manufacturers, importers, and distributors?

A business label alone does not decide which licence applies. A company described commercially as a manufacturer, importer, or distributor may perform more than one regulated activity in Canada. Consider the device class, the named manufacturer, each Canadian establishment’s activities, and applicable exemptions.

Business role or activity Potential MDL requirement Potential MDEL requirement What to verify
Manufacturer Yes, for a Class II-IV device it manufactures for sale in Canada Assess separately based on establishment activities and exemptions Device class, manufacturer identity, Canadian activities, applicable exemptions
Importer The device may require an MDL held by its manufacturer if it is Class II-IV Potentially, for importing devices, subject to exemptions Device licence status, importing activity, exemption criteria
Distributor The device may require an MDL held by its manufacturer if it is Class II-IV Potentially, for distributing devices, subject to exemptions Device licence status, distribution activity, exemption criteria

For a manufacturer, start with the device. If the manufacturer intends to sell a Class II-IV device in Canada, that device requires an MDL before sale. Separately, for an importer or distributor, start with the establishment’s activities. The product’s MDL status remains a separate question.

This activity-based approach helps avoid two common errors:

  • Treating an MDEL as though it were a product licence.
  • Assuming that an existing MDL eliminates the need to assess establishment activities.

When does a medical device need an MDL?

The MDL requirement applies to Class II, III, and IV devices. Class I devices do not require an MDL under section 26 of the Medical Devices Regulations.

Before deciding whether a device needs an MDL:

  1. Identify the device and its intended use.
  2. Classify the device using Schedule 1 of the Medical Devices Regulations.
  3. Confirm the manufacturer named for the device.
  4. For a Class II-IV device, confirm whether an MDL is required before sale in Canada.

Health Canada’s Medical device licences information is a useful starting point for checking licence-related information. Resolve classification before making a licence decision, especially where the intended use is uncertain.

When does an establishment need an MDEL?

Section 44 of the Medical Devices Regulations establishes the requirement to hold a Medical Device Establishment Licence (MDEL) for importing or selling medical devices, subject to the applicable exemptions.

The practical question is what the Canadian establishment actually does, not simply whether the business calls itself an importer or distributor. For each establishment, document whether it imports devices, distributes devices, or performs another activity relevant to the MDEL assessment.

Review Health Canada’s Medical device establishment licences page alongside the regulation when completing that assessment.

How can you determine which licence your business needs?

Use a documented, device-by-device and establishment-by-establishment review. This is more reliable than reaching one conclusion for the entire corporate group or product portfolio.

  1. Create a device inventory. List each device offered in Canada, its Canadian classification, its intended use, and its named manufacturer.
  2. Review MDL status. Identify Class II–IV devices and verify whether each device has an active Medical Device Licence (MDL) held by the manufacturer using Health Canada’s Medical Devices Active Licence Listing (MDALL).
  3. Map Canadian establishments. Identify each Canadian location or entity involved in the supply chain.
  4. List actual activities. Record whether each establishment imports, distributes, or performs other relevant activities.
  5. Assess MDEL requirements. Compare each establishment’s activities against the applicable requirements and exemptions under the Medical Devices Regulations, including sections 44 and 45, and relevant Health Canada guidance.
  6. Document the conclusion. Retain the basis for each decision, including any exemption relied upon.
  7. Reassess after changes. Recheck the analysis when the business changes activities, supply chain arrangements, or device portfolio.

A written assessment records why the business concluded that an MDL, MDEL, both licences, or neither applies. It also makes later review easier if device classifications, manufacturer arrangements, or Canadian distribution activities change.