ISO 13485 Production and Service Controls

Validation engineer reviewing IQ, OQ and PQ protocol results beside a medical device sealing machine in a controlled manufacturing area

Clause 7.5 of ISO 13485:2016 governs how a medical device is made, labeled, cleaned, sterilized, installed, serviced, identified and protected until it reaches the user. It is also where process validation, including IQ, OQ and PQ for medical devices, becomes a formal requirement.

Since February 2, 2026, FDA's Quality Management System Regulation (QMSR) has applied clause 7.5 in the US, with added labeling, packaging and UDI record requirements. These controls start at  at design transfer and run for as long as a device is made, distributed and serviced.

Scaling up production or preparing for an audit?

Production, Installation and Servicing

 Control of Production (7.5.1)

Production must be planned, carried out, monitored and controlled. In practice that means documented procedures and methods, qualified infrastructure, monitored process parameters, defined product release, controlled labeling and packaging operations, and a record for each device or batch showing the quantity made and approved for distribution.

 Installation and Servicing Activities (7.5.3–7.5.4)

In the case of an installation being required, acceptability criteria should be set and installation and verification data recorded. When a third party undertakes the installation task, provide him with documented requirements. In cases where servicing is a mandatory requirement, use documented procedures and reference measurements, and analyze servicing records.

Cleaning, Identification and Traceability

 Cleanliness of Product (7.5.2)

Cleanliness specifications are required when cleaning of the product is necessary before sterilizing or using it, when the product is supplied non-sterile, the product needs to remain uncleaned even though cleaning is needed, or when process materials need removal.

Identification and Traceability (7.5.8–7.5.9)

Identify product and its inspection status throughout production, segregate returned product, and apply a UDI where regulations require one. Traceability procedures set the records needed; implantable devices also need records of components, materials and work environment conditions, plus distribution records.

Sterile Device and Sterilization Process Controls

Clause 7.5.5 requires sterilization parameters to be recorded for each sterilization batch and traced to each production batch. Clause 7.5.7 requires sterilization processes and sterile barrier systems to be validated before use and after changes. ISO 11135 (ethylene oxide), ISO 11137 (radiation), ISO 17665 (moist heat) and ISO 11607 (packaging for terminally sterilized devices) are the usual basis, whether you sterilize in-house or through a contract sterilizer.

Parul Chansoria, Founder and CEO of Elexes, on ISO 13485 production and service controls, validated processes and device traceability

Process and Software Validation

Gowned cleanroom operator scanning a batch label at a laminar-flow workstation beside a partially assembled medical device and component trays.

When a Process Needs Validation (7.5.6)

Validate any procedure in which the output does not lend itself to, or is not capable of, later measurement or testing, such as sealing, welding, molding or sterilization. Procedures encompass approval criteria, equipment and personnel qualification, methodology, statistical methodologies and rationale for sample size, record keeping, and revalidation.

 IQ, OQ and PQ for Medical Devices

ISO 13485 doesn't name IQ, OQ and PQ, but most manufacturers validate in these stages. Installation qualification confirms equipment is installed to specification, operational qualification establishes process limits, and performance qualification shows the process consistently produces acceptable product under normal conditions. Link the results to risk management.

Validation of Production and Quality System Software

Any software that is used for manufacturing or delivering services should undergo validation before its first use and following any modifications, in proportion to the risk involved. The Computer Software Assurance guidance document issued by FDA in September 2025 provides a risk-based approach to such software.

Product Preservation

Clause 7.5.11 requires conformity to be preserved through processing, storage, handling and distribution, using suitable packaging and shipping containers or documented special conditions such as temperature limits. Clause 7.5.10 adds that customer property, including intellectual property and confidential health information, is identified, verified, protected and safeguarded.

Production and Service Control Expectations by Market

Under the QMSR, FDA adds 21 CFR 820.45 labeling and packaging controls, and 820.35 requires the UDI to be recorded for each device or batch and in servicing records. The EU MDR sets UDI obligations in Article 27, with manufacturing and sterilization covered in technical documentation. Production and service controls is a core MDSAP audit area for Health Canada, and the MHRA and TGA expect the same controls within the QMS.

How Elexes Supports ISO 13485 Production and Service Controls

Elexes helps manufacturers, contract manufacturers and specification developers turn clause 7.5 into procedures, protocols and records that hold up in audits. We assess current practice, prioritize gaps by risk, prepare documentation and support FDA, EU MDR, UK, Health Canada and TGA audits, as part of our ISO 13485 consulting services. We work alongside your test labs and sterilization providers rather than replacing their testing.

Our Production and Service Control Services

Work instructions, release criteria, labeling controls and records for each device or batch.

Validation master plans, protocols, sample-size rationale and report review.

Risk-based validation for production and quality system software.

Planning and reviewing validation with your sterilizer and packaging suppliers.

Identification, traceability and UDI records across production, distribution and servicing.

Closing findings and preparing teams for FDA, notified body and MDSAP audits.

Assess Your ISO 13485 Production and Service Controls Readiness

Our checklist covers production, cleanliness, traceability, sterilization, validation and preservation, so you can find gaps before an auditor does.
FAQs

Frequently Asked Questions!

ISO 13485 requires validation of processes whose output isn’t fully verified afterwards, but it doesn’t mandate the IQ, OQ and PQ labels. They remain the most common way to structure and document that validation.

Yes. According to clause 7.5.6, the production software needs to be validated before initial use and changes and depending on the risks involved in using it.

Since February 2, 2026, FDA applies ISO 13485 clause 7.5 in place of the former 21 CFR 820.70 and 820.75, while keeping US-specific labeling, packaging and UDI record requirements.

A Trusted Team of Consultants

Elexes offers a globally recognized ISO 13485 certification consultant who is renowned to enhancing your business credibility in the medical devices industry.

Local Language
Auditor

Our presence across the globe ensures audits are conducted by local-language auditors, facilitating clear communication and understanding.

Why Choose Elexes?

Expertise

Our experts provide guidance and consultants to help you adhere to necessary regulations, ensuring your products meet all legal requirements and is compliant to all necessary standards.

Local Regulatory Certification

At Elexes, we offer to assist in securing additional local certifications, enabling you to manufacture and sell medical devices in various markets across the globe.

Confidentiality

We offer complete confidentiality to all our clients which makes us one of the most trusted regulatory and quality compliance consulting.

Associated Regulatory Authorities.

img
We’ve More
Than 256+ Global Clients
The Best RA & QA Consultants
Clients Testimonials

Amazing Feedback About Our Services

Popular Clients

We’ve 200+ Global Premium Clients

Saneso
Niramai
Apex dental sensors
Neurasignal
SS inovations
Amaryllis
Inclode
iMedisync
DJOGlobal logo cmyk_TM
Summus Laser Logo
Looking For Regulatory Assistance?

Let's Talk How We Can Help You?

Please enable JavaScript in your browser to complete this form.
Name
img
Please enable JavaScript in your browser to complete this form.
Name
img
Please enable JavaScript in your browser to complete this form.
Name
img
Please enable JavaScript in your browser to complete this form.
Name
img
Please enable JavaScript in your browser to complete this form.
Name
img