IVDD vs IVDR: the IVDR replaced the IVDD and changed IVD classification, conformity assessment, performance evidence, technical documentation, and post-market obligations. Manufacturers should assess each device against the IVDR and confirm whether amended transition provisions apply.
This comparison is for IVD manufacturers placing devices on the EU market, especially teams evaluating new products, implementing changes, or reviewing the regulatory status of legacy IVDD devices.
What is the difference between IVDD and IVDR?
The IVDD was Directive 98/79/EC. The IVDR is Regulation (EU) 2017/746, which establishes the EU framework for in vitro diagnostic medical devices and repealed the IVDD through Article 112 of the regulation. See the official Regulation (EU) 2017/746.
A central practical difference is the legal form. An EU regulation is directly applicable in Member States, while a directive requires implementation through national law. Previous IVDD conformity or certification should not be treated as automatic evidence that a device meets IVDR requirements. The European Commission provides an overview of the IVDR framework for in vitro diagnostic medical devices.
For manufacturers, the IVDD versus IVDR question is not simply whether a certificate remains available. It requires a device-specific review of intended purpose, classification, conformity-assessment route, evidence, documentation, quality-system controls, post-market processes, and any transition eligibility.
| Topic | IVDD | IVDR |
|---|---|---|
| Legal instrument | Directive 98/79/EC | Regulation (EU) 2017/746 |
| Classification approach | Categories and lists, including Annex II lists | Risk-based Classes A, B, C, and D under Annex VIII |
| Conformity assessment | Route depended on the applicable IVDD category and route | Route depends on the IVDR class and applicable provisions in Article 48 and relevant annexes |
| Performance evidence | Requirements under the IVDD | Performance evaluation addresses scientific validity, analytical performance, and clinical performance |
| Transition provisions | Legacy devices were subject to applicable IVDD provisions | Check Article 110 as amended, including current eligibility conditions and device-specific dates |
How did IVD classification change under the IVDR?
The IVDD used categories and lists, including Annex II lists and self-testing devices. The IVDR uses risk-based Classes A, B, C, and D. The applicable class is determined using the classification rules in Annex VIII.
This change means an IVDD category should not simply be carried forward as an IVDR class. The review should begin with the device’s current intended purpose and then apply the relevant Annex VIII rules.
Classification matters because it informs the conformity-assessment route and the level of notified-body involvement. However, a class alone does not answer every assessment question. The applicable route also depends on the device and its characteristics under Article 48 and the relevant annexes.
A classification reassessment should document:
- The intended-purpose statement being assessed
- The applicable Annex VIII classification rule or rules
- The resulting IVDR class
- The rationale for the classification conclusion
- Any assumptions or open questions requiring further review
What changed for conformity assessment and notified bodies?
Under the IVDR, the conformity-assessment procedure depends on the device class and the applicable route in Article 48 and Annexes IX to XI. Specified IVDR routes require notified-body involvement.
Class A devices that are not sterile generally follow a different route from devices subject to notified-body assessment. Manufacturers should therefore confirm the exact route for the device rather than making a decision based only on a broad class label or an apparently similar product.
Before building an assessment plan, review whether the intended notified body is designated for the relevant IVDR scope and conformity-assessment activities. The European Commission’s notified bodies information page is an appropriate starting point for that review.
Practical questions to resolve include:
- Which IVDR classification rule applies?
- Which conformity-assessment route applies to that class and device characteristic?
- Is notified-body involvement required for the selected route?
- Is the selected notified body designated for the relevant IVDR scope and activities?
- What documentation and evidence will support the selected route?
Teams planning a certification pathway can also consider IVD certification support when assessing the route and related documentation needs.
What evidence and documentation should manufacturers reassess?
The IVDR requires a performance evaluation that addresses scientific validity, analytical performance, and clinical performance. Manufacturers should assess whether their available evidence, and the resulting performance evaluation report, support the device’s intended purpose.
Technical documentation should also be reviewed against the IVDR requirements in Annexes II and III. This includes the device description and specification, design and manufacturing information, and performance-evaluation documentation.
The review should extend beyond the technical file. Article 10 sets out manufacturer obligations, while the IVDR also includes requirements relevant to post-market surveillance and vigilance. Quality-management-system and post-market processes should therefore be checked for applicable IVDR requirements.
Key documentation and process areas to reassess include:
- Device description, specification, and intended purpose
- Design and manufacturing information
- Scientific validity, analytical performance, and clinical performance evidence
- The performance evaluation report
- Technical documentation controls
- Post-market surveillance processes
- Vigilance processes
For organizations reviewing their operational controls alongside technical documentation, post-market surveillance for IVDs can help frame the post-market workstream.
How should you perform an IVDD-to-IVDR gap assessment?
An effective gap assessment is device-specific. It should identify what applies to each product, what evidence and documentation are available, and what actions are needed to support the applicable IVDR pathway.
| Area to review | Questions for the manufacturer | Potential follow-up |
|---|---|---|
| Device inventory and intended purpose | Is each device and its intended purpose clearly documented? | Confirm the device scope and intended-purpose statements |
| Classification | What IVDR class follows from the intended purpose and Annex VIII rules? | Document the classification rationale and review applicable rules |
| Conformity assessment | Which IVDR route applies, and is notified-body involvement required? | Confirm the route and assess notified-body planning needs |
| Performance evaluation and technical documentation | Do evidence and technical documentation address applicable IVDR requirements? | Identify evidence gaps and update controlled documentation |
| Quality system and post-market processes | Do processes address applicable IVDR and device-specific obligations? | Review and remediate quality-system and post-market procedures |
| Transition status | Does the device meet the current eligibility conditions for any applicable transition? | Verify Article 110 as amended and document the device-specific assessment |
A practical process is:
- Build a device inventory. Record each device’s intended purpose, current classification, conformity-assessment status, certificates, and current market status.
- Reassess IVDR classification. Apply the Annex VIII rules to the intended purpose and document the rationale.
- Determine the conformity-assessment route. Map the device to the applicable IVDR route and identify whether notified-body involvement is required.
- Assess evidence and technical documentation. Review performance evaluation, technical documentation, and controlled records against the applicable IVDR requirements.
- Review quality-system and post-market controls. Check whether relevant processes address IVDR manufacturer obligations, post-market surveillance, and vigilance.
- Confirm transition status where relevant. For a device relying on transition provisions, check current Article 110 requirements as amended by Regulation (EU) 2024/1860. Confirm device-specific eligibility conditions and applicable dates before relying on a transition.
- Prioritize and track remediation. Record each gap, its rationale, an owner, and the next action. Prioritize gaps affecting the conformity-assessment route, evidence adequacy, technical documentation, or continued transition eligibility.
Frequently asked questions about IVDD vs IVDR
What is the difference between IVDD and IVDR?
IVDD is Directive 98/79/EC, while IVDR is Regulation (EU) 2017/746. The IVDR changed the classification framework and conformity assessment, and it requires performance evaluation, technical documentation, and post-market processes to be assessed against the applicable IVDR requirements.
How do I transition from IVDD to IVDR?
Assess each device’s intended purpose, IVDR classification, conformity-assessment route, performance evidence, technical documentation, and post-market processes. If the device relies on a transition, verify the current Article 110 provisions as amended by Regulation (EU) 2024/1860, including device-specific conditions and dates.
Get support with an IVDD-to-IVDR assessment
A focused IVDR gap assessment can help establish the applicable classification, conformity-assessment route, evidence needs, documentation gaps, and transition status for each device. Need help assessing your IVD’s IVDR gaps or transition requirements? Talk to Elexes about IVDR consulting for device-specific gap assessments.