An EU MDR post-market clinical follow-up (PMCF) plan should define device-specific objectives, planned activities, methods, procedures, and the rationale for proactively collecting and evaluating clinical data. The plan should address outstanding clinical questions and residual risks in accordance with Annex XIV, Part B of Regulation (EU) 2017/745.

This guide explains how manufacturers can select appropriate PMCF activities, document their rationale, and structure the plan using MDCG 2020-7. It also explains the relationship between historical MEDDEV 2.12/2 guidance and current EU MDR requirements.

What should an EU MDR PMCF plan cover?

A PMCF plan should explain what clinical evidence is needed for the specific device, how that evidence will be proactively collected and evaluated, and how the findings will support the ongoing clinical evaluation and risk management process.

Under Regulation (EU) 2017/745, Annex XIV Part B, the PMCF plan must specify appropriate methods, procedures, objectives, rationale, relevant references, and a justified schedule for PMCF activities.

At a practical level, the plan should document:

  • Device identification and intended purpose.
  • PMCF objectives and the clinical questions the activities are intended to answer.
  • Methods and procedures for proactively collecting and evaluating clinical data.
  • The rationale for each selected method.
  • How data will support confirmation of safety and performance throughout the expected lifetime of the device.
  • How data will support assessment of the continuing acceptability of identified risks and detection of emerging risks.
  • How results will feed into the clinical evaluation and risk management processes.

The MDCG 2020-7 PMCF Plan Template provides a practical structure for recording this information. It is guidance, not legislation, but it can help teams check whether the plan presents a coherent approach to evidence generation and assessment.

Which specific PMCF activities can manufacturers consider?

The right activity depends on the clinical question. A PMCF plan may contain one activity or several complementary activities, provided each is tied to a defined evidence need and an intended contribution to the clinical evaluation.

Activity Potential clinical question Plan details to define
PMCF study or clinical investigation Does the device perform as expected in the intended population or use setting? Objective, study method, population, data collection and analysis approach
Registry or structured post-market data source What clinical outcomes or safety signals are observed in routine use? Registry or source, relevance to the device, data fields and evaluation method
Targeted patient or user follow-up Are there unresolved questions about outcomes, use or longer-term experience? Who will be followed, how information will be collected and what question it addresses
Literature review and clinical-experience analysis Does published evidence or relevant clinical experience add information about safety and performance? Search and selection approach, relevance criteria and analysis method

When should you consider a PMCF study or clinical investigation?

A PMCF study or other clinical investigation may be appropriate when existing clinical evidence leaves an unresolved question requiring prospective clinical data. The plan should state the specific question the investigation is intended to answer.

Define the study method, target population, data-collection approach and analysis method in relation to that question. Listing a study without explaining the evidence gap it addresses does not show why it is part of the device-specific PMCF strategy.

When can registries or targeted follow-up address the question?

A relevant device or disease registry, structured post-market data source, or targeted patient or user follow-up may be appropriate where it can address a defined question about the device in routine use.

The plan should identify:

  • The registry, data source or follow-up population.
  • Why the source is relevant to the device under evaluation.
  • The data fields or information to be assessed.
  • The method for evaluating the resulting data.
  • The clinical question the activity is intended to address.

When should you include literature review and clinical-experience analysis?

Systematic literature review and analysis of clinical experience may contribute to PMCF when they can provide relevant information on the device’s safety or performance. Experience with comparable devices can also be considered where the comparison is justified and relevant to the device under evaluation.

Describe the search and selection approach, relevance criteria and analysis method. The plan should also state how the findings will contribute to the clinical evaluation, rather than treating the review as a standalone administrative task.

For a broader implementation perspective, see Elexes’ practical PMCF implementation guidance.

How should you choose PMCF activities for a particular device?

Start with unresolved evidence questions, not a standard catalogue of activity types. Under Annex XIV Part B, PMCF is a proactive and systematic process for collecting and evaluating clinical data to confirm safety and performance, ensure the continuing acceptability of identified risks and detect emerging risks.

Use this process to select activities:

  1. Identify unresolved clinical-evaluation questions. Review the clinical evaluation for areas where additional clinical evidence is needed.
  2. Review residual risks. Identify risks that require continuing clinical confirmation or further data assessment.
  3. Consider the device context. Take account of the device’s characteristics, intended purpose, patient population, expected use and relevant post-market information.
  4. Match each question to an activity. Select a study, registry, targeted follow-up, literature review or other method based on its ability to address the identified uncertainty.
  5. Document the rationale. Explain why the method is suitable and proportionate to the question being investigated.
  6. Record material exclusions. Where a decision not to conduct an activity needs explanation, document the rationale.
  7. Define the intended use of evidence. State how the data will be evaluated and incorporated into the clinical evaluation and risk management processes.

A generic activity list is not a device-specific PMCF plan. The plan should connect the evidence gap or residual risk to the selected activity, data assessment approach and resulting clinical-evidence output.

What should a PMCF plan specify for each activity?

For every planned activity, record enough detail to support consistent execution and meaningful evaluation. Specify its objective, design or method, data source or population, and approach to assessing the resulting data.

Use this cross-check when reviewing a draft PMCF plan:

Plan element What to document Related clinical-evidence output
Device and objectives Device identification, intended purpose and the clinical questions or objectives for follow-up Defined evidence needs for the clinical evaluation
Activities and rationale Selected methods, data sources or populations, and why they address the identified questions Planned collection of relevant clinical data
Evaluation approach How collected data will be assessed and interpreted PMCF evaluation report and updated clinical evaluation
Use of findings How findings will inform clinical evaluation and risk management Relevant clinical and post-market documentation

In particular, each activity should include:

  • Its objective and clinical question.
  • The design or method.
  • The data source, population or information set.
  • The approach for evaluating the data.
  • Assigned responsibilities.
  • How results will be documented.
  • How findings will feed into clinical evaluation and risk management.

The anticipated outputs should be consistent with the MDCG 2020-8 PMCF Evaluation Report Template. Reviewing the expected evaluation-report structure while drafting the plan can help align the planned data, evaluation method and intended conclusions.

PMCF findings should support the continuing clinical evaluation rather than remain isolated in a post-market document. See Elexes’ guide to how PMCF findings support the clinical evaluation report.

How does MEDDEV 2.12/2 relate to EU MDR PMCF requirements?

MEDDEV 2.12/2 can provide useful context for PMCF planning, but it should not be treated as a standalone statement of current legal requirements. For EU MDR devices, Annex XIV Part B of Regulation (EU) 2017/745 provides the binding PMCF requirement baseline.

A practical review approach is to:

  1. Use EU MDR Annex XIV Part B as the binding requirement baseline.
  2. Use MDCG 2020-7 as a practical reference for PMCF plan structure.
  3. Use MDCG 2020-8 to assess whether planned activities can generate a coherent PMCF evaluation report.
  4. Confirm the current status and wording of guidance before relying on a specific recommendation.

The European Commission’s medical-device guidance document library should be checked for the current availability and status of MDCG documents and other guidance.